Subscribe to Test & Measurement World
RSS
Reprints/License
Print
Email
Average Rating:
  • (4)
    Rate this:
  • Thermocouples often described incorrectly

    Martin Rowe, Senior Technical Editor -- Test & Measurement World, 10/1/2008 2:00:00 AM

    For reactions to this article, see Martin Rowe's blog posting, "Thermocouples raise reader temperatures."

    Perhaps the most common mistake told about measurement is the phrase “A voltage is developed across the junction of a thermocouple that’s proportional to temperature.” It’s repeated so often that even experienced engineers take it for granted. It’s also incorrect.



    Thermocouples generate a difference voltage proportional to a temperature difference.

    A thermocouple is made two of wires of different metals where the tips are connected (often welded) to produce a good electrical connection. This junction is the temperature measurement point. Figure a shows the incorrect assumption about thermocouples. There is no voltage developed across the junction, although there will be a voltage across the open ends of the wires (VT).

    In Figure b, assume that the thermocouple junction is at temperature T1 and, for simplicity, assume that the open ends of the wires are both at T0 = 0°C. A wire whose ends are at different temperatures will produce a small voltage across those ends, called the Seebeck Effect. That voltage depends on the composition of the wire and the temperature difference across the wire’s ends.

    The material of a wire affects the voltage across its ends, so wires of two different materials produce two different voltages (V1 and V2) for the same temperature difference. Thus, the voltage across the terminals is VT = V1 – V2. The voltage develops across the wires, not across the junction.

    Because keeping the wire ends at 0°C is impractical, equipment manufacturers use a circuit to measure the temperature at the ends of the thermocouple wires and electronically compensate for that temperature. It’s called cold-junction compensation (CJC). Even with CJC, the thermocouple wire connections must be as close to each other’s temperature as possible to minimize measurement errors. Most connections to thermocouple measuring equipment, therefore, are mounted on a thermally conductive block.

    Average Rating:
  • (4)
    Rate this:
  • RSS
    Reprints/License
    Print
    Email
    Similar Content from T&MW

    No related content found.

    »MORE

    • 0 rated items found.

    Datasheets.com Electronic Parts & Inventory Search

    185 million searchable parts
    • Part Number
    • Description
    • Inventory
    • Products
    • Manufacturers
    Canon Resource Center

    Featured Company


    Most Recent Resources

    Featured Job On
    Scroll for More Jobs
    Advertisement
    More Content
    • Blogs
    • Webcasts

    Sorry, no blogs are active for this topic.

    » VIEW ALL BLOGS RSS
    • All


    Advertisement
    Advertisement
    About Us   |   Advertising Info   |   Site Map   |   Contact Us   |   FREE Subscription
    © 2011 UBM Electronics . All rights reserved.
    Use of this Web site is subject to its Terms of Use | Privacy Policy

    Feedback Form
    Feedback Analytics