Build a Magnetic Field Immunity Tester
A precompliance test system can help you determine whether your products comply with standards such as the CE Marking.
Jim Ericson, Acme Testing, Acme, WA -- Test & Measurement World, 6/1/1999
| The latest versions of European CE-Marking
standards include magnetic-immunity requirements for residential, commercial, industrial
and audio/visual electrical and electronic equipment. Your product must be immune to
magnetic fields from transformers or other sources. These fields can produce character
jitter on CRT displays, distortion in audio equipment, or false readings in
equipment containing Hall-effect devices or other electromagnetic field sensors. You can assemble a precompliance (or even full compliance) magnetic field immunity test system in a few hours using common laboratory instruments, a 100-W audio amplifier, some white PVC Schedule-40 3/4-in. water pipe, some wood, and a copper or aluminum sheet. European Standard EN 61000-4-8:1993 requires a 63 dB test field tolerance.1 That requirement results in a usable test volume of 0.6 x 0.6 x 0.5 m for a square loop 1 m on a side. Larger EUTs would require construction of a larger loop or a double Helmholtz coil as shown in the test standard. The Biot-Savart Law states that a single-turn circular loop with a diameter of 1-m will generate a magnetic field at its geometric center equal to the loop current. So, a round 1-m diameter loop carrying 1 A of current with two turns will produce a field of 2 A/m at its center. For a square loop, the correction factor is 4/(p=2), or 90%. A square 1-m loop with two turns carrying 1 A of current will, therefore, generate a 1.8 A/m field at its center. Magnetic fields, however, are usually expressed in gauss, teslas, and oersteds. See Handy Conversions, this page, for help in converting units. Build the Loop
Make all connections mechanically strong. Solder them well because the total loop resistance at 50 Hz will be about 200 mV. To complete the test setup, build a wood frame to support the loop. Place the frame and loop above a non-magnetic, aluminum or copper ground plane with minimum dimensions of 1 m x 1 m x 0.25 mm. Figure 2 shows how to connect the loop to the audio amplifier through an 8-W, 100-W power resistor in series with the loop. The resistor approximates the impedance of a speaker. The amplifier must be capable of having the low side of its output connected to safety earth. Connect the low side of the loop coil circuit to the ground plane; connect the ground plane to safety earth.
You can also measure the loop current. Place a current probe or a DMM (set for ACA) in series with the loop. Verify the meters accuracy at the frequencies you plan to use. You also need to verify that the loop current sinusoidal distortion is less than 8%. Do that with a distortion analyzer or visually by monitoring the loop voltage waveform with an oscilloscope as you adjust the signal input level. When you begin to overdrive the amplifier and distortion of the sine wave reaches about 5%, the distortion will start to be visible on the scope pattern. A 100-W audio amplifier will supply about 2.5 A into an 8-W load without appreciable distortion. This
corresponds to a maximum magnetic field at the center of the 2-turn 1-m square loop of 2
turns x 2.5 A x 0.9 turns/m, or 4.5 A/m. If you need higher field intensities, either
increase the number of loop turns or drive the loop directly from the AC mains through a
variac and stepdown transformer.
Test Procedure After verifying the EUTs performance to its technical specifications, expose the EUT (and 1 m of any EUT cables) to the test field at different orientations by rotating the loop and shifting the EUT. Verify that the effects of the magnetic field on the EUT are acceptable and that the equipment is able to recover by itself when you remove the field. You can use the loop to perform full compliance testing if you calibrate the loop. To perform the calibration, you must measure the field strength with a calibrated instrument. Measure the field generated by the magnetic loop with either a Gaussmeter or a small calibration loop antenna.2 If you use the loop antenna, you must connect it to either an EMI meter or a spectrum analyzer. Depending upon the EUTs application, the requirements of the generic light and heavy industrial standards EN 50082-1:1997 and EN 50082-2:1995 for magnetic immunity may range from 3 A/m to 30 A/m at the powerline frequency.3,4 The product family standard for EMC immunity of professional
audio/video equipment (EN 55103-2:1996) requires magnetic immunity testing from 50 Hz to
10 kHz at levels from 0.008 A/m to 10.0 A/m, depending upon the EUTs
classification.5 Requirements for other families (such as medical devices) are under
consideration. T&MW FOOTNOTES Jim Ericson is an EMC engineer at Acme Testing. acmetest@acmetesting.com. |






















