Global TMW:
Login  |  Register          Free Newsletter Subscription
Subscribe
Email
Print
Reprint
Learn RSS

Counter/timer accurately calibrates frequency

A high-resolution counter/timer with an external frequency reference lets you calibrate frequency to an uncertainty of 10–12 or better.

Staffan Johansson, Pendulum Instruments, Stockholm, Sweden -- Test & Measurement World, 5/1/2002

Accuracy versus resolution in counter/timers

An accurate way to calibrate a frequency source is to use a high-resolution counter/timer to measure the beat frequency, or phase difference, between the source to be calibrated and a reference such as a cesium or a GPS-controlled rubidium standard. The method measures how fast the phase difference between the unknown source and the reference increases or decreases. One requirement of this method is that the source and the reference have the same nominal frequency.

Figure 1. A counter/timer can measure TIE based on the zero-crossing times of both the unknown frequency and the reference frequency.
Figure 2. Measuring TIE at two points lets you calculate the frequency difference between two signals with the same nominal frequency (Df/f = DTIE/Dt).


The easiest way to determine this phase difference is to measure the time interval, or time interval error (TIE), between zero crossings of the two signals with a counter/timer (Figure 1). By measuring the TIE at two points several seconds apart, you can calculate a relative frequency difference (Äf/f), as shown in Figure 2:

Äf/f =
TIE2 – TIE1/t 2t 1

To illustrate, I measured Äf for an unknown 10-MHz signal with moderate short-term frequency fluctuations using an extremely stable 10-MHz signal from a Fluke 910R, a GPS-controlled frequency reference with a built-in rubidium oscillator. I also used a Fluke PM6681 counter/timer with 50-ps resolution to measure the TIE every 20 s. The results are shown in Table 1.

Table 1. Counter/timer measurement results
Time t (s) TIE (ns) ÄTIE (ps) Ät (s) ÄTIE/Ät = Äf/f
t0 = 0 +4.55
t1 = 20 +4.75 200 20 1.0 x 10–11
t2 = 40 +4.99 440 40 1.1 x 10–11
t3 = 60 +5.23 680 60 1.1 x 10–11
t4 = 80 +5.49 940 80 1.2 x 10–11
t5 = 100 +5.72 1170 100 1.2 x 10–11

You can see that a measurement taken over 100 s shows that the two frequencies differ by 1.2 x 10–11. You should note, though, that even the 20-s reading gives a good indication of the difference. Knowing the difference, you can calibrate the unknown source either by documenting its value or by adjusting it to reduce the difference if that facility exists.

No calibration is complete without knowing the measurement uncertainty. So, what's the uncertainty of this measurement? There are four steps to attaining measurement uncertainty:

  1. Compensate the result for any known systematic errors.
  2. Express the remaining uncertainty factors as standard deviations (such as rms or 1-ó values).
  3. Take the square root of the sum of the squares of all uncertainty contributions to get a standard uncertainty.
  4. Multiply the standard uncertainty by 2 to reduce the risk of a measurement value being outside the area of uncertainty. For a normal distribution, 2-ó means 96% of all measurements will be within the given area of uncertainty.
  • In my example, the counter/timer measures TIE with 50-ps resolution and measures t with 100-ns resolution. The relative frequency difference is ÄTIE/Ät. The TIE uncertainty at t 0 and t 5 is resolution (50 ps) plus the same systematic error (t syst) in both TIE measurements. Because you subtract TIE(t 0) from TIE(t 5), the calculation eliminates the systematic error (t syst), and only the uncertainty of the resolution remains in ÄTIE.

    The uncertainty of the time difference ÄT = t 5t 0 (100 ns in 100 s) is negligible. Therefore, the total measurement uncertainty becomes:

    The relative frequency difference (Äf/f) is thus (1.2 ±0.14) x 10-11 in the example.

    By measuring ÄTIE over longer periods (Ät), you can improve resolution and reduce the measurement uncertainty even more. Other sources of error in the measurement, such as the counter/timer's timebase oscillator stability and trigger errors due to noise, are negligible compared to the resolution of the TIE measurement.

    Note that the calculation in the example gives the measurement uncertainty of the difference between the two frequencies. In order to determine the accuracy of the frequency being calibrated, you have to consider the accuracy of the reference itself. For a Fluke 910R, deviations appear in the 12th digit. For cesium standards, deviations appear in the 13th digit.

    Direct calibration is simpler

    If you don't need such a high-accuracy method for calibrating frequency, you can simply connect the signal you want to calibrate to an already-calibrated frequency counter and read the result. The big advantage of this method is that you can calibrate any frequency within the range of the counter. The disadvantage of using a direct reading is a limit of approximately 10 digits of measurement accuracy even if you use the highest-resolution counters with a built-in rubidium reference. The frequency measurement time—averaging period—is normally limited to 10 s.

    To find the measurement uncertainty of direct frequency calibration, you must add internal-timebase error, trigger error due to signal noise and internal noise, and systematic timing error in the counting circuits. Noise errors normally show up in the 9th or 10th digit of a reading, systematic errors usually occur in the 10th or 11th digit.

    As an example, if both resolution and timebase uncertainty are 1 x 10–7 rms, and the other sources of errors are negligible, then the total uncertainty will be:

    Today's counters have a typical resolution of 9 or 10 digits (in some cases up to 11 digits) for a measuring time of 1 s. It's the built-in timebase oscillator that limits the accuracy (see "Accuracy versus resolution in counter/timers," p. 24).

    Theoretically, you could use an external cesium standard or a GPS-controlled rubidium reference to obtain 11- to 12-digit measurement uncertainty, but in practice, trigger noise and systematic errors in the counter usually limit any improvement to 9 to 10 digits.

    Even with the best frequency counters, you won't beat 10-digit measurement uncertainty using direct-frequency calibration. To achieve anything better, you must use the phase comparison method.

    Note: A version of this article previously appeared in Test & Measurement Europe.



    Author Information
    Staffan Johansson is marketing manager with Pendulum Instruments, Stockholm, Sweden, and has 20 years experience in the test and measurement business. He has presented many seminars and lectures on time and frequency analysis, and he has written books on measurement techniques.

  •  

    Accuracy versus resolution in counter/timers

    A built-in timebase oscillator in a counter/timer requires regular calibration, typically once a year, to compensate for the oscillator's long-term drift. Variations of ambient temperature also affect the oscillator frequency.

    Counter/timers may display up to 10 to 12 digits, but the validity of these digits depends largely on the type of internal timebase oscillator the counter/timer has. Typical values for use in normal room temperature and with 1-yr calibration interval are:

    • Uncompensated crystal oscillator: 5–6 reliable digits
    • Temperature-compensated crystal oscillator: 6–7 reliable digits
    • Oven-controlled crystal oscillator: 7–9 reliable digits
    • Rubidium oscillator: 10–11 reliable digits

    Using an external cesium standard or GPS-controlled reference with a counter/timer can produce even higher stability. The GPS-controlled reference avoids aging because the built-in local oscillator locks to the cesium standards in GPS satellites. The locking process, however, can reduce short-term stability, especially for low-grade local oscillators.

    Figure A. A GPS-controlled rubidium oscillator gives a level of accuracy approaching that of a cesium oscillator.
    The frequency stability can be 1 x 10–12 over 24 hrs. When locking a local-oven oscillator, the stability can be 100 to 1000 times worse over short periods (1 to 10 s), which are typical measuring periods in frequency counters. If you lock a local rubidium oscillator, the locking process is not measurable, and you have a frequency reference comparable to the cesium reference at a fraction of the cost (Figure A).

    The resolution of a counter depends upon its measurement technique, which can be conventional, reciprocal, or interpolating.

    Conventional frequency counting is no longer used in frequency counter instruments. With this method, instruments count the number of cycles during a fixed period, usually 1 s. The resolution of the measurement is then 1 cycle/s, or 1 Hz. The display resolution then depends on what frequency you measure. For 100 Hz, the resolution is 1 Hz/100 Hz or "2 digits" (10–2). For 10 MHz, the relative resolution is 1 Hz/10 MHz or "7 digits" (10–7).

    Reciprocal counters measure the period of the input signal by counting clock pulses (nCP) from a 10-MHz reference clock for an integer number of periods (nP) of the input signal. Usually the counters measure over a period of 1 s. After that, you can calculate the period (t):

    t=nCP /nP x 100 ns

    and the frequency (f), is the reciprocal of t:

    f = 1/t = nP /nCP x 10 MHz

    The resolution with this method is always 7 digits for a 1-s measuring period (100 ns resolution in 1 s) and independent of the frequency.

    Figure B. Relative resolution for different frequency counting techniques.
    Interpolating counters are reciprocal counters but with improved time measurement resolution. This type of instrument not only counts the integer number of clock pulses given by the time measurement, but it also calculates by interpolation the incomplete parts of clock pulses at the beginning and at the end of the time measurement. This technique improves resolution by 100 to 400 times, and it makes time measurements with a resolution of hundreds of picoseconds.

    Independent of counting method, resolution always improves with increased measuring time. Normally, counters can measure over 10 s, which provides one digit of resolution more than a measurement made over a 1-s period (see Figure B).

  • Email
    Print
    Reprint
    Learn RSS

    Talkback

    We would love your feedback!

    Post a comment

    » VIEW ALL TALKBACK THREADS

    Related Content

    Related Content

     

    By This Author

    There are no other articles written by this author.

    Sponsored Links



     
    Advertisement
    SPONSORED LINKS

    More Content

    • Blogs
    • Podcasts

    Blogs

    • Rick Nelson
      Taking the Measure

      July 1, 2008
      S-parameters are so yesterday
      Textbook amplifiers operate in linear mode and are easy to analyze. Unfortunately, it’s often ...
      More
    • Rick Nelson
      Taking the Measure

      June 30, 2008
      Cell phones helping cell phones
      Now, I’m leery of the phrase “paradigm shift,” which is often applied to increment...
      More
    • » VIEW ALL BLOGS RSS

    Podcasts

    Advertisements





    NEWSLETTERS
    Click on a title below to learn more.

    Test Industry News (3 Times Per Month)
    Machine-Vision & Inspection (Monthly)
    Communications Test (Monthly)
    Design, Test & Yield (Monthly)
    Automotive, Aerospace & Defense (Monthly)
    Instrumentation (Monthly)
    Resource Center E-Alert (Monthly)
    ©2008 Reed Business Information, a division of Reed Elsevier Inc. All rights reserved.
    Use of this Web site is subject to its Terms of Use | Privacy Policy
    Please visit these other Reed Business sites