Global TMW:
Login  |  Register          Free Newsletter Subscription
Subscribe
Email
Print
Reprint
Learn RSS

More than just point & shoot

IR thermometers and cameras need accurate temperature sources and good calibration practices to achieve optimal results.

Martin Rowe, Senior Technical Editor -- Test & Measurement World, 12/1/2003

Everything emits

Infrared (IR) thermometers and imaging cameras can locate thermal hot spots on ICs, assembled circuit boards, and mechanical devices without physical contact. Like any measuring instrument, IR thermometers (also called "pyrometers") and cameras perform only as well as their calibration. Blackbody temperature sources let you calibrate IR temperature instruments and minimize errors.

An imaging-camera calibration lab uses numerous black-body calibrators. Courtesy of FLIR Systems.

Noncontact thermometry instruments detect IR radiation—which is proportional to temperature—from their targets. To calibrate an IR thermometer or camera, you need a standard source at a known temperature that emits a known level of radiation. Because blackbodies emit the highest level of IR radiation for a given temperature, they're the preferred standard for calibrating IR temperature instruments. (See "Everything emits ,".)

A blackbody calibration source often contains a thermal cavity with an aperture opening that provides the IR thermometer or camera with access to the cavity (Figure 1). A heating or cooling unit, a temperature probe (usually a thermocouple), and a temperature controller monitor and control the cavity's temperature. Calibrator temperatures can reach 2000°C.

No cavities

Not all blackbody temperature calibrators contain cavities. Some use a thermal plate (Figure 2). Blackbody plate calibrators let you calibrate low-end IR thermometers and perform checks of IR cameras prior to calibration.

IR imaging camera manufacturers often start with a plate calibrator. A blackbody plate calibrator provides a consistent temperature across the place from which a technician verifies that all of a camera's sensors produce the same color on a computer screen. With the plate, a technician can adjust the camera to nullify offsets that produce errors.

Once a camera produces a consistent color, it's ready for a full calibration. At FLIR Systems (Billerica, MA; www.flirthermography.com), the calibration lab consists of several blackbody calibrators arranged in a quarter circle. The camera sits on a small platform at one end of the arm; the other end connects to a vertical rotating pole at the center of the circle. Two wheels that support the table end of the arm let technicians rotate the camera to point at any blackbody calibrator.

The calibrator temperatures range from –196°C to 2000°C, says service manager Pat McDonough. A monitor displays the image that the camera captures. A technician takes readings at each temperature and enters them into the camera. A curve-fit algorithm generates a calibration look-up table of digital values versus temperature for each of the camera's calibration ranges.

Ircon (Niles, IL; www.ircon.com) takes a different approach. Instead of using numerous blackbody calibrators, Ircon uses one. Technicians calibrate IR imaging cameras at five temperatures, says VP Vern Lappe. This process takes longer than using multiple calibrators, but it costs less to maintain.

Ircon's calibration procedure includes temperature cycling of the camera. To perform a calibration, a technician places the camera in a thermal chamber that cycles from 0°C to 45°C for each of the blackbody calibrator's temperature settings. The camera's thermal chamber contains an opening for the infrared sensors.

Placement counts

Regardless of whether you use one or many blackbody calibrators, you need to properly place the IR instruments relative to each calibrator. To perform an accurate calibration, you must place the instrument so its measuring area (field of view) is smaller than the area of the target (the calibrator's plate size or aperture size). Ideally, the target area should be twice that of the field of view, because an instrument takes the average temperature within its field of view and the temperature differs from the center to the edges.

Figure 3. Calibration requires placing the thermometer or camera at the proper distance from the calibrator.

Figure 3 shows three locations for positioning relative to the blackbody calibrator. Location A marks the best place for performing a calibration. Location B works, but not necessarily as well, because of the larger measuring area. If the calibrator's temperature varies in the measured area, it will affect the measured value. At location C, the thermometer's field of view exceeds the blackbody calibrator's aperture size. Thus, the measured value will appear lower than the blackbody source's temperature. When calibrating an IR instrument, place the instrument as close to the calibrator as possible, but maintain a safe distance from high temperatures.

Ircon's Lappe offers more tips for calibrating IR thermometers.

  • Always place the thermometer on a solid stand and aim it at the cavity.
  • Check the thermometer's calibration as soon as it comes into the lab. If it is out of calibration and the factory uses ISO procedures, notify the quality staff about the instrument's condition.
  • Clean the instrument's lens according to the manufacturer's instructions. Some thermometers have three lenses to clean. Remember, IR thermometers and imaging cameras make optical measurements, so a clean lens improves accuracy.
  • After cleaning, aim the instrument at the blackbody calibrator and check the temperature. Check the lower, middle, and upper ranges of the instrument.
  • During the calibration, make sure the instrument is completely warmed and the blackbody calibrator's temperature is stable.
  • Check the thermocouples used in blackbody sources at least once a year because thermal cycling of a blackbody source can, over time, damage its thermocouple. At temperatures such as 2000°F, the cavity glows (Figure 4). Thermal cycling causes expansion and contraction of the thermocouple's materials, which can degrade the sensor.
Calibrator calibration

Because a calibration is only as good as the calibrator, a blackbody calibrator itself needs calibration. High-accuracy IR thermometers, with calibrations traceable to NIST or another national lab, serve as transfer standards and usually carry uncertainties four times better than the blackbody sources they calibrate.

You can also calibrate a blackbody calibrator with a spectroradiometer, says Robert Madding, director of training at FLIR Systems' Infrared Training Center (www.infraredtraining.com). A spectroradiometer measures the IR power spectrum of its target, producing a plot of IR power as a function of wavelength. "Spectroradiometric calibration is the only way to truly calibrate the irradiance from a blackbody source," claims Madding.

Blackbody calibrators provide manufacturers and third-party calibration labs with accurate temperature sources for IR thermometers and imaging cameras. Most users send imaging cameras to the manufacturer for calibration. Proper care and good calibration practices will produce instruments that consistently meet manufacturers' accuracy specifications.

For more information

"IR Thermometer Calibration," Omega Engineering, Stamford, CT. www.omega.com/literature/transactions/volume1/calibrate1.html.

Madding, Robert P., "Spectroradiometric Calibration of Blackbody Sources," Proceedings of Thermosense (XXIII), Vol. 4360. International Society for Optical Engineering (SPIE), Bellevue, WA, April 2001. p. 363. www.photonics.org/web/abstracts/4300/4360.html.

For more information about camera topics, visit www.tmworld.com/ins.


Author Information
Merchant, John, "Blackbody calibration sources function as standards," Application note BB, Mikron Instruments, Oakland, NJ, 1995. www.mikroninst.com/App_Notes/apnoteBB.pdf
"The Theoretical Basis For Radiation Measurements," Mikron Instruments, 2002. www.mikroninst.com/products/blackbody/about-blackbody.html.

 

Everything emits

Every object at temperatures above absolute zero (0 K) absorbs energy and emits electromagnetic radiation. If an object absorbs all the energy incident upon it, it's called "black." Think of it as a black hole in space, which absorbs everything within its gravitational pull. A perfect blackbody absorbs all the energy around it and reflects none.

Objects also emit IR radiation in proportion to the fourth power of their temperature, according to the Stefan-Boltzmann Law (Ref. 1). The amount of radiated energy, called irradiance or radiant exitance, is expressed in watts per square centimeter, W/cm2 (Ref. 2). A blackbody is also a perfect emitter of IR radiation. That is, no object can emit more IR radiation at a given temperature than a blackbody (Ref. 3).

The ratio of emitted energy to reflected energy is a material's emissivity. A perfect blackbody, therefore, has an emissivity of 1. A perfect reflector (emissivity = 0) neither absorbs nor emits energy, but rather, it reflects energy. Although no material is a perfect emitter, some objects and organic material come close. "The human body comes close to perfect emissivity," says Scott Saborin, applications engineer at Isotech North America (Burlington, VT; www.isotechna.com), but you can't calibrate an instrument to a person.

Blackbody sources contain heated areas coated with materials whose emissivity exceeds 0.9 (Ref. 4). A blackbody cavity, because of its modified conical shape, produces an effective emissivity of 0.995 or greater. A plate blackbody source often contains a groove in its heating plate that brings its effective emissivcity to 0.95 or greater. Both types of calibrators maintain their emissivity over the commonly used wavelengths of 3–5 µm and 8–14 µm. —Martin Rowe

REFERENCES

  •  Stefan-Boltzmann Law," Georgia State University, Atlanta, GA. ewhdbks.mugu.navy.mil/EO-IR.htm
  • "Radiation Terms," Electro Optical Industries, Santa Barbara, CA. www.electro-optical.com/bb_rad/define.htm
  • "Material Emissivity Properties," Electro Optical Industries, Santa Barbara, CA. www.electro-optical.com/bb_rad/emissivity/matlemisivty.htm

    1. Email
      Print
      Reprint
      Learn RSS

      Talkback

      We would love your feedback!

      Post a comment

      » VIEW ALL TALKBACK THREADS

      Related Content

      Related Content

       

      By This Author

      Sponsored Links



       
      Advertisement
      SPONSORED LINKS

      More Content

      • Blogs
      • Podcasts

      Blogs

      • Rick Nelson
        Taking the Measure

        July 1, 2008
        S-parameters are so yesterday
        Textbook amplifiers operate in linear mode and are easy to analyze. Unfortunately, it’s often ...
        More
      • Rick Nelson
        Taking the Measure

        June 25, 2008
        CEOs address proposed Credence, LTX integration
        Credence and LTX complement each other with respect to customers, product lines, facilities, and emp...
        More
      • » VIEW ALL BLOGS RSS

      Podcasts

      Advertisements





      NEWSLETTERS
      Click on a title below to learn more.

      Test Industry News (3 Times Per Month)
      Machine-Vision & Inspection (Monthly)
      Communications Test (Monthly)
      Design, Test & Yield (Monthly)
      Automotive, Aerospace & Defense (Monthly)
      Instrumentation (Monthly)
      Resource Center E-Alert (Monthly)
      ©2008 Reed Business Information, a division of Reed Elsevier Inc. All rights reserved.
      Use of this Web site is subject to its Terms of Use | Privacy Policy
      Please visit these other Reed Business sites